CVE-2019-25722EPSS p9.1%
CVE-2019-25722CVE-2019-25722
Description
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
Scoring
| CVSS | 7.6 () |
| Vector | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H |
| EPSS | 0.19% probability of exploitation · percentile 9.1% · 2026-06-19T12:03:05Z |
| Last modified | 2026-06-04 |
Related by meaning· 6