CVE-2019-25719EPSS p3.0%

CVE-2019-25719CVE-2019-25719

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Scoring

CVSS 8.6 ()
VectorCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
EPSS0.13% probability of exploitation · percentile 3.0% · 2026-06-19T12:03:05Z
Last modified2026-06-03

Related by meaning· 6

Nearest entities by semantic similarity across the cs-graph corpus.

CVE
CVE-2019-25724
CVE
CVE-2019-25721
CVE
CVE-2022-4992
CVE
CVE-2019-25716
CVE
CVE-2019-25717
CVE
CVE-2019-25720
Sourced from NVD + FIRST.org EPSS. Curated for EU compliance use cases by Adam Lundqvist, Founder at SQUR.