CVE-2025-14096HIGH 8.4EPSS p1.9%
CVE-2025-14096CVE-2025-14096
Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.
Other related CVE's are CVE-2025-14095 & CVE-2025-14097.
Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure:
Attacker requires physical access to the analyzer.
Temporary work Around:
Only authorized people can physically access the analyzer.
Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Scoring
| CVSS 3.1 | 8.4 (HIGH) |
| Vector | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| EPSS | 0.12% probability of exploitation · percentile 1.9% · 2026-06-21T12:00:28Z |
| Published | 2025-12-17 |
| Last modified | 2026-04-15 |
Underlying weaknesses· 2
References
2
| Type | Target | Confidence | Tier |
|---|---|---|---|
| Weakness | Execution with Unnecessary Privilegescwe-250 | 0% | live |
| Weakness | Use of Hard-coded Credentialscwe-798 | 0% | live |
Related by meaning· 6
Nearest entities by semantic similarity across the cs-graph corpus.