CVE-2022-43557EPSS p12.1%

CVE-2022-43557CVE-2022-43557

bd / bodyguard_999-603_firmware

Description

The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.

Scoring

CVSS 5.3 ()
VectorCVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H
EPSS0.22% probability of exploitation · percentile 12.1% · 2026-06-19T12:03:05Z
Last modified2026-06-17
Sourced from NVD + FIRST.org EPSS. Curated for EU compliance use cases by Adam Lundqvist, Founder at SQUR.